The Flame Retardant Tester Comply with UL1581 standards,It is suitable for wire and cable insulation flame resistance test;
1. Scope of application:
It is suitable for wire and cable insulation flame resistance test;
Vertical test: place the specimen on the top of the flame, burn for 15 seconds, extinguish for 15 seconds, repeat 5 times and check the burning condition of the specimen. The burning, extinguishing time and repetition times can be set, and automatic operation can be performed.
2. Compliant standards:
2.1 Comply with UL1581-2016 Section 1061 Vertical Burning Test of Wires and UL1581-2016 Section 1090 Horizontal Burning Test of Equipment Wires;
2.2 Comply with the test of Section 9.2 of UL2556-2015;
3. Main parameters:
3.1 Vertical combustion box: made according to UL1581 standard size, the internal size is 305*355*610mm.
3.2 Horizontal combustion box: made according to UL1581 standard size, the internal size is 305*355*610mm.
3.3 Vertical spark nozzle: The nozzle angle is 20 degrees, with gas control valve.
3.4 Horizontal spark nozzle: The nozzle angle is 90 degrees, with gas control valve.
3.5 Vertical or horizontal nozzle selection mode.
3.6 Select manual/automatic mode.
3.7 When the preset data is reached, the machine will automatically stop the test.
3.8 Fuel: Gas. Methane (provided by the customer)
3.9 Test length: 300mm (horizontal), 457mm (vertical).